The Purpose of Using Quartz Cuvettes for Spectrometer and their Importance
The
apparatus used for the measurement of absorbance is known as spectrophotometer.
Beam of light can be produced by the
spectrophotometer in the wavelength given. The device requires tools for
conducting experiments on samples. The
tools include a sample holder or a Cuvette holder a light source, a monochromator,
a display and a detector. Depending upon
the wavelength hydrogen with high pressure and xenon lamp with high pressure or
deuterium is used as the light source. The purpose of the monochromator is to choose
light from the light source spectrum for the wavelength given. Samples holder enables the light to pass
through the sample. Sample holder or cuvette holders made up of glass, plastic
and quartz.
Quartz Cuvettes – An Overview
Hellma Quartz Cuvettes are used for spectrometer based on need and demand. These apparatus
use quartz Cuvettes when the incident light reaches the UV range. Basically
quartz Cuvettes will not absorb light while on the other hand plastic or glass
Cuvettes absorbs light which will in turn produce wrong results. Quartz
Cuvettes are also used for samples containing solutions in order to avoid absorption
of light. Quartz Cuvettes are transparent to UV range and also provides better
response in the UV range. Plastic and glass Cuvettes do not have these
properties so they are not used in the spectrophotometer. Properly cleaned and
maintained with care, quartz Cuvettes can be reused.
For
conducting any measurements the condition and quality of the Cuvette is important.
By using thermostatted holder the samples are maintained at constant
temperature throughout the measurement. While measuring absorbance at a wavelength of 320
mm quartz Cuvettes are used. Glass and plastic will absorb light in the range
of 320 mm. The length of the path in
most Cuvettes will be 1 cm and the volume is 1.5 ml. While conducting
measurements on absorbance narrowed Cuvettes with darkened sides are
recommended.
Spectrophotometer uses single as well as double beams to measure the sample. In case of single
beam only one sample is measured at a time. While the apparatus holding double beam
conducts measurement on two samples where one Cuvette holder holds the standard
sample and the other holds the reference sample. Electronic detector which is
light sensitive is used to measure the light intensity.
Using
spectrometers absorbance can be measured for given wavelength or for the
spectra as a whole by changing the wavelengths. The present values of the
spectra can be taken by the spetrophotometer which is up to date.

Comments
Post a Comment